Applications 115
5.5.2
DACONLY
1.
Press the CURSOR
menu soft key
and a small circle will appear on the display that selects
the points/echo where DAC compensation is calculated and the amplitude of the selected
points is adjusted. With the UP/DOWN arrow keys or the SmartKnob™, move the cursor
to the point/echo where the DAC Only is to start (see
Figure 5-56
).
2.
Press the SELECT
menu soft key;
this will log the next point/echo. SELECT will
automatically change to ADJUST. With the UP/DOWN arrow keys or SmartKnob™,
adjust the gain to increase the echo amplitude to the desired height: for example, from
70% FSH (full screen height) to 80% FSH (see
Figure 5-57
).
3.
Press the CURSOR
menu soft key
, and with the UP/DOWN arrow keys or SmartKnob™,
adjust the cursor to the next point/echo (see
Figure 5-58
).
4.
Press the SELECT menu soft key; this will log the next point/echo. SELECT will
automatically change to ADJUST. With the UP/DOWN arrow keys or SmartKnob™,
adjust the gain to increase the echo amplitude to the desired height: for example, from
60% FSH to 80% FSH (see
Figure 5-59
).
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...