Applications 65
5. Applications
What's in this section?
•
5.1 Pulse-Echo Contact Thickness Testing
•
5.2 Dual Transducer Thickness Testing
•
5.3 Delay Line Thickness Testing
•
5.4 Shear Wave (Angle Beam) Testing
•
5.5 Distance Amplitude Correction (DAC) Testing
5.1
Pulse-Echo Contact Thickness Testing
One of the major methods of ultrasonic thickness testing is the pulse-echo contact method. It
provides excellent accuracy and is best suited for smooth surfaces. The test range is usually
0.10 inch to 20.0 inches. A wideband transducer is normally used. The thickness gauge offsets
and sound velocity must be recalibrated each time the instrument is powered up.
The following accessories are required for this setup procedure:
STEP BLOCK
— Sonic
®
#1920105.00 Steel Step Block or equivalent 0.100,
0.200, 0.300, 0.400, and 0.500 inch steps.
TRANSDUCER
— The transducer used will depend on the material and
thickness being tested. For steel or aluminum up to 5 inches thick, a
Spectrum
E8L
(5 MHz) transducer is a good selection. For rough, corroded, or highly
attenuative materials, a 2.25 MHz transducer
(Spectrum™ E4L)
may be required.
Larger and lower frequency transducers may give better results on rough
materials (that is, cast iron).
TRANSDUCER CABLE
— Use a Sonic
®
BNC/Microdot (#9102894) or
BNC/BNC (#9102892) to connect the specific transducer.
Содержание Sonic 1200S/HR
Страница 1: ...PN 7720044 PN 7720066 March 2006 Sonic 1200S HR Ultrasonic Flaw Detector User s Manual...
Страница 8: ......
Страница 13: ...Preparation for Operation 5 1 2 Sonic 1200S HR Packages 1 2 1 Sonic 1200S Packages...
Страница 14: ...6 Chapter 1...
Страница 15: ...Preparation for Operation 7 1 2 2 Sonic 1200HR Packages...
Страница 16: ...8 Chapter 1 1 3 Sonic 1200S HR Packages 1 3 1 Sonic 1200S Packages...
Страница 17: ...Preparation for Operation 9...
Страница 18: ...10 Chapter 1 1 3 2 Sonic 1200HR Packages...
Страница 36: ...28 Chapter 2...
Страница 72: ...64 Chapter 4...
Страница 110: ...102 Chapter 5 Figure 5 37 IIW Type 1 Block Echoes Figure 5 38 IIW Type 2 Block Echoes...
Страница 119: ...Applications 111 f Calibration is now done see Figure 5 50 Figure 5 48 Figure 5 49...
Страница 134: ...126 Chapter 6...
Страница 149: ...Computer Interface 141...