8.5 Performance test
8-5
FRA 51602
3) Testing equipment
The measurement instruments needed for checking operation and testing performance are listed as below.
Product
Required specifications
Example models
Frequency counter
Accuracy: ±1 ppm or better
53131A
+ option 001
(Keysight)
Digital multimeter
AC voltage (true rms)
(100 mV to 10 V, 100 Hz to 10 kHz)
Accuracy: ±0.1%
DC voltage
Accuracy: ±0.1%
(100 mV to 10 V)
34401A
(Keysight)
RF voltmeter
AC voltage (true rms)
(100 mV to 10 V)
Accuracy: ±0.5%
(1 kHz to 100 kHz)
Accuracy: ±1.0%
(100 kHz to 1 MHz)
Accuracy: ±5.0%
(1 MHz to 2 MHz)
URE3
(Rohde & Schwarz)
Other
BNC-BNC coaxial cable 50 Ω, RG-58A/U
× 2
BNC 50 Ω feed-thru terminator
× 1
(Power rating is 0.25 W or more)
*1
BNC-banana adapter
× 1
BNC adapter (provided)
× 1
BNC short plug
× 1
BNC-alligator clip cable
× 1
*1: Not required if the input impedance of the RF voltmeter is 50 Ω.
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...