3.5 Operation screen reference
3
-
62
FRA 51602
■ [SEARCH VALUE]
–
[X]
: X search value
■ [SEARCH VALUE]
–
[Y1]
: Y1 search value
■ [SEARCH VALUE]
–
[Y2]
: Y2 search value
■ [SEARCH VALUE]
–
[
⊿
X]
:
⊿
X search value
■ [SEARCH VALUE]
–
[
⊿
Y1]
:
⊿
Y1 search value
■ [SEARCH VALUE]
–
[
⊿
Y2]
:
⊿
Y2 search value
This sets the value to be searched by marker.
○ Setting method
< Numerical value settings (immediate setting) > , < Numerical value settings (new input) > ,
< Numerical value settings (edit) >
○ Function key functions
• [DIRECT NEW]
: Set the maximum value or minimum value of the X axis with the numerical
input dialog
【
NUM
】
.
• [DIRECT EDIT]
: Edit the maximum value or minimum value of the X axis with the
numerical input dialog
【
NUM
】
.
○ Functions of the function keys set with the numerical input dialog
【
NUM
】
• [u(10
–6
)]
: Set the entered value in units of 10
–6
.
• [m(10
–3
)]
: Set the entered value in units of 10
–3
.
• [×1]
: Set the entered value.
• [k(10
3
)]
: Set the entered value in units of 10
3
.
• [M(10
6
)]
: Set the entered value in units of 10
6
.
• [f(10
–15
)]
: Set the entered value in units of 10
–15
.
• [p(10
–12
)]
: Set the entered value in units of 10
–12
.
• [n(10
–9
)]
: Set the entered value in units of 10
–9
.
• [G(10
9
)]
: Set the entered value in units of 10
9
.
• [T(10
12
)]
: Set the entered value in units of 10
12
.
The setting range is –1.0 T to +1.0 T (10
12)
and the resolution is 6 digits or 1 f (10
–15
), whichever is
larger. When the search is for frequency, the range for the main marker is from 10 µHz to 2 MHz and
the resolution is 10 µHz. For the frequency delta marker, the range is from –2.999,999,999,990 MHz
(–2 MHz + 10 µHz) to +2.999,999,999,990 MHz (2 MHz –10 µHz) and the resolution is 10 µHz.
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...