8.5 Performance test
8-6
FRA 51602
8.5.2 Oscillator frequency accuracy
Before performing each of the following measurements, initialize this instrument.
This instrument
: Set the frequency to 1.0 MHz.
Measuring instrument
: Use a frequency counter and set the gate time to 10 seconds.
Connections
: Connect the
OUTPUT OSC
connector of this instrument to the
frequency counter with a coaxial cable.
Measurement method
: Read the measured value from the frequency counter.
Judgment
Passing range
: 0.999990 kHz to 1.000010 MHz
Measured value
: _ ._ _ _ _ _ _ MHz
FRA 51602
OSC
CH1
CH2
BNC-BNC
coaxial cable
Frequency counter
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...