5.5 Graph display
5
-
17
FRA 51602
■ The unwrapped phase display function
For the
[UNWRAP]
display setting, a wrapped phase display is performed, with the phase measurement
value at the sweep starting frequency (the lower-limit frequency for an upsweep or the upper-limit
frequency for a downsweep) as the reference value. Thus, when DUTs that have entirely the same phase
characteristics are measured by upsweep or downsweep measurement, the results are displayed in graphs
that differ in phase by an integer multiple of 360°.
■ The 360° phase shift function
To display the currently-displayed phase data as a graph with 360° added or subtracted, select
【
Graph
】
–
[PHASE RANGE]
–
[SHIFT]
and tap on either of the function keys labeled
[+360°]
or
[
−360°]
. This is
used when a phase value in a phase continuous display is too large. When the phase display range setting
is anything other than
[UNWRAP]
, the display is not changed.
2) Tap
1) Select SHIFT
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...