3.2 Input and output connectors
3
-
18
FRA 51602
■ 10 MHz REF IN
If a clock signal is input to this connector, that replaces the internal reference clock of the instrument and
the instrument will operate on the external clock. The accuracy of the measurement frequency is
determined by the accuracy of the clock frequency that is input. Frequency-synchronized measurement
across multiple devices is possible by using a common reference clock.
Input voltage
: 0.5 Vp-p to 5 Vp-p
Input waveform
: Sinusoidal or square
Input impedance
: 300 Ω (nominal), AC coupling
Maximum non- destructive voltage
: 10 Vp-p
Signal ground
: Isolated from enclosure, ±42 V (DC + ACpk) max.
Connector
: Rear panel, BNC connector
!
WARNING
Do not apply voltages that exceed the rated isolation voltage. Doing so may result in
insulation breakdown and risk of electrical shock or damage to the instrument.
1
MΩ
10 MHz
REF IN
EXT REF
CKT
3
3
0
k
Ω
(3
0
0
Ω
)
±
42 V
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...