3.5 Operation screen reference
3
-
71
FRA 51602
■ [WEIGHTING FACTOR]
–
[CH1]
: Set the CH1 weight factor.
■ [WEIGHTING FACTOR]
–
[CH2]
: Set the CH2 weight factor.
This value serves as a multiplier for the measurement result. It can be used as a simple correction factor
for when a probe, amplifier, or shunt resistance is connected to the circuit being measured.
○ Setting method
< Numerical value settings (immediate setting) > , < Numerical value settings (new input) > ,
< Numerical value settings (edit) >
○ Function key functions
• [DIRECT NEW]
: Set the maximum or minimum X axis value with the numerical input
dialog
【
NUM
】
.
• [DIRECT EDIT]
: Edit the maximum or minimum X axis value with the numerical input
dialog
【
NUM
】
.
○ Functions of the function keys set with the numerical input dialog
【
NUM
】
• [u(10
– 6
)]
: Set with the entered value in units of 10
– 6
.
• [m(10
– 3
)]
: Set with the entered value in units of 10
– 3
.
• [×1]
: Set with the entered value.
• [k(10
3
)]
: Set with the entered value in units of 10
3
.
• [M(10
6
)]
: Set with the entered value in units of 10
6
.
• [p(10
– 12
)]
: Set with the entered value in units of 10
– 12
.
• [n(10
– 9
)]
: Set with the entered value in units of 10
– 9
.
• [G(10
9
)]
: Set with the entered value in units of 10
9
.
• [T(10
12
)]
: Set with the entered value in units of 10
12
.
The setting range is 0.0 to 1.0 T (10
12
) and the resolution is 6 digits or 1 p(10
– 12
), whichever is larger.
■ [WEIGHTING FACTOR]
–
[INVERT]
: Set phase inversion.
This inverts the phase of the measurement results.
○ Setting method
< List operations (confirming) >
• ON
: Display the measurement results with phase inversion.
• OFF
: Do not invert phase.
For more information, refer to "
5.9.2 Input weighting
".
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...