3.2 Input and output connectors
3
-
17
FRA 51602
■ VGA
This connector outputs a video signal that is the same as the instrument LCD to an external monitor or
projector.
Connector
: 15-pin mini-D-sub (female), analog RGB component video signal
Signal
: 800 × 600 pixel (SVGA)
Horizontal frequency = 37.9 kHz; vertical frequency = 60.3 Hz
■
DC BIAS OUT
This is an output connector for supplying DC bias to the equipment under test.
Maximum output voltage : ±10 V
Output impedance
: 600 Ω ±2%
Connector
: Rear panel, BNC connector
If the connector to which the DC bius voltage is output is set to the rear-panel
DC BIAS OUT
connector,
the DC bias (DC voltage) is output from this connector. In that case, only the AC component is output
from the
OSC
connector on the front panel.
About the setting of the DC bias
→
"
4.3 Oscillator settings
".
The output impedance is 600 ohms, regardless of whether the output is on or off. Even when the output is
set to off, the relay is constantly connected to the internal output amplifier (the relay is always on). For
that reason, turning the output on or off does not affect the output impedance.
When the power supply is off, the relay is also off and the output amplifier is disconnected from the
rear-panel
DC BIAS OUT
connector. Even in that state, the output impedance remains at 600 ohms.
However, when the power is turned on or turned off, the output impedance is temporarily unstable.
The output voltage is set when there is no load (open load). When a 600 ohm load is connected, the
output is at half of the set voltage.
!
CAUTION
・
Applying an external signal to the DC BIAS OUT connector will damage the internal
circuits. Never input a signal to the DC BIAS OUT connector.
・
The BNC ground of the DC BIAS OUT connector is connected to the enclosure. Do
not connect signals for which there is a difference in potential relative to the
enclosure. Doing so may damage the instrument.
600
Ω
DC BIAS OUT
Enclosure
Relay
Output Amp
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...