2.3 Grounding and power connections
2
-
11
FRA 51602
2.3 Grounding and power connections
■ The instrument must be grounded.
!
WARNING
This instrument uses line filters and there is risk of electrical shock if the instrument
is not grounded.
To prevent electrical shock, connect it to an earth ground so that ground resistance is
100
Ω or less.
This instrument is automatically grounded when the provided power cord with a three-terminal plug is
inserted into three-terminal ground-protected outlet.
■ Power supply specifications
Power supply voltage range
: 100 VAC to 230 VAC ±10%, not to exceed 250 V
Frequency
: 50 Hz / 60 Hz ±2Hz
Power consumption
: 100 VA, maximum
Overvoltage category
: II
■ Connecting the power supply
1) Confirm that the voltage of the mains to which the instrument is connected is within the specified
voltage range of the instrument.
2) Set the power switch that is on the rear panel to the off position.
3) Insert the power cord into the power inlet that is on the rear panel of the instrument.
4) Plug the power cord into a ground-protected power outlet.
5) Set the power switch to the on position and keep the power switch on during normal operation of the
instrument.
!
CAUTION
The power cord set that comes with the instrument is intended for use only with this
instrument.
The power cord set should not be used with any other product or for any other
purpose.
When connecting the instrument to the mains, be sure to use only the power cord
that is provided with the instrument.
If the power to the instrument is turned off, wait at least one minute before turning it back on again.
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...