5.4 Oscillator control
5
-
11
FRA 51602
5.4.4 Measurement synchronization
This function automatically turns the oscillator output off when the measurement is completed. The setting
is done with
【
OSC
】
–
[ON/OFF SYNC]
. This function is used when it is not desirable for the measurement
signal to be applied to the equipment under test for a long period of time.
[ASYNC]
: The oscillator output is not turned off, even when the measurement is completed.
[SYNC]
: The oscillator output is turned off when the measurement is completed.
The order of operations for when the setting is SYNC is shown below.
Start measurement (Press the
UP
,
DOWN
, or
SPOT
key.)
↓
Oscillator output is turned on.
↓
Wait for the START DELAY time. (Measurement is not performed during this time.)
↓
Start signal acquisition (begin analysis)
↓
(frequency changes during the sweep)
:
End analysis (the sweep ends)
↓
Oscillator output is turned off
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...