5.9 Error correction
5
-
37
FRA 51602
5) When the sweep measurement is completed, select
【
Trace
】
–
[MEAS COPY]
and tap the
[EQUALIZE CORR MEM]
function key to copy the measurement data to the internal equalizing
memory of the instrument.
6) Set
【
Calibration
】
–
[EQUALIZING]
to
[ON]
.
7) Restore the original connections as they were prior to removing the EUT in step 3 ("
Figure 5-18
Example of connections for gain-phase measurement
").
The procedure described above completes the equalizing preparation. Performing a sweep measurement
produces measurement data that is corrected for error in probes, cables and other such parts of the
measurement system.
The correction data that has been copied to the equalizing memory is retained, even if the instrument power
is turned off. It is therefore available and ready for use the next time the power is turned on. However, this
function is executed as part of the error correction process for the instrument itself, so we recommend that
the equalizing correction data be measured again when the internal temperature of the instrument has
stabilized after the power is turned on (about 30 minutes) or when there has been a sudden change in the
ambient temperature, etc.
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...