7.1 Error messages
7
-
4
FRA 51602
Table 7-2
Panel operation errors (continued)
No.
Error message
Description and cause
Handling
-6145 There is already a file of
the same name.
A file of the same name already exists,
so the specified name cannot be used to
rename the file.
Specify a different file
name.
-6146 USB storage overflow.
There isn't sufficient available space on
the USB memory device.
Insert a USB memory
device that has sufficient
available space.
-6147 File load error due to
illegal header.
The file could not be loaded because it
is an invalid file (invalid header).
Specify a valid file.
-6148 File load error due to
illegal data.
The file could not be loaded because it
contains invalid data.
Specify a correctly
formatted file.
-6149 Failed to save.
The process of saving the file to the
USB memory device failed.
Remove the read-only
attribute from the file.
-6150 Failed to load.
The process of reading the file from the
USB memory device failed.
Specify an undamaged
file.
-6151 Failed to delete.
The process of deleting the file from
the USB memory device failed.
Remove the read-only
attribute from the file.
-6152 Can't load empty
memory.
The process of reading the
measurement data failed because the
internal memory is empty.
Specify in internal
memory that is not
empty.
7.1.3 Errors that occur during measurement
The errors that may occur during measurement are listed in
Table 7-3
.
Table 7-3
Errors that occur during measurement
No.
Error message
Description and cause
Handling
-3072 Measurement terminated
due to measurement
points over 20,001.
The number of measurement points
exceeded the limit of 20,001, so
measurement was stopped.
When performing sequence
measurement, reduce the
number of measurement
points at each step.
When performing automatic
high density sweep
measurements, change the
threshold setting.
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...