3.5 Operation screen reference
3
-
40
FRA 51602
3.5.1 OSC dialog
■ [FREQUENCY]
: Set the frequency
This is the frequency of the signal that is output from the
OSC
connector. The frequency at which a spot
measurement is performed (measurement at a fixed frequency) is set. After a sweep measurement,
however, the value is replaced by the last measured frequency. If the sweep is interrupted, the frequency
at the time of interruption is set.
○ Setting method
< Numerical value settings (immediate setting) > , < Numerical value settings (new input) > ,
< Numerical value settings (edit) >
○ Function key functions
・
[SWEEP MIN
– > FREQ]
: Copy the sweep lower frequency limit to
FREQUENCY
.
・
[SWEEP MAX
– > FREQ]
: Copy the sweep upper frequency limit to
FREQUENCY
.
・
[MKR
– > FREQ]
: Copy the main marker frequency to
FREQUENCY
.
・
[
⊿
MKR
– > FREQ]
: Copy the delta marker frequency to
FREQUENCY
.
・
[DIRECT NEW]
: Set the frequency with the numerical input dialog
【
NUM
】
.
・
[DIRECT EDIT]
: Edit the frequency setting with the numerical input dialog
【
NUM
】
.
○ Functions of the function keys set with the numerical input dialog
【
NUM
】
・
[uHz]
: Set with the entered value in units of µHz.
・
[mHz]
: Set with the entered value in units of mHz.
・
[Hz]
: Set with the entered value in units of Hz.
・
[kHz]
: Set with the entered value in units of kHz.
・
[MHz]
: Set with the entered value in units of MHz.
The setting range is 10 µHz to 2 MHz; The setting resolution is 10 µHz.
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...