FRA 51602
!
WARNING
Do not apply voltages that exceed the rated isolation voltage. Doing so may result in
insulation breakdown and risk of electrical shock. Damage to the instrument may also
result.
The maximum input voltage is restricted according to the frequency as described below. Applying high
voltages at high frequencies may result in a large AC current flowing in floating capacitance within the
instrument, creating a risk of damage to the equipment. Be certain to use the instrument at voltages below
the rated isolation voltage.
○ Maximum input voltage by frequency
Below 45 Hz
: 600 Vpk (AC+DC)
45 Hz to 100 kHz
: 600 Vrms
100 kHz to 2 MHz
: 600 Vrms (AC) × 100 ÷ F, where F is frequency in kHz
For frequencies of 45 Hz or higher, the RMS voltage including DC is 600 Vrms or less and the peak
voltage is 849 Vpk or less.
Frequency vs Maximum Input Voltage (CH1,CH2)
10
100
1k
1
10
100
1k
10k
100k
1M
10M
Frequency [Hz]
M
a
x
im
u
m
I
n
p
u
t
V
o
lt
a
g
e
[
V
]
AC Vrms
(AC+DC)
Vpk
DC
~
Figure 3-5
Maximum input voltage of the analysis module input connectors by frequency
This rating is for when there is no transient overvoltage. For measurements in commercial mains circuits
were entrancing over voltages can occur, use the MEASUREMENT CATEGORY specifications
described previously.
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...