6.3 Measurement conditions file format
6
-
7
FRA 51602
16
AmplitudeCompressionRefCh, {1|2}
Sets the amplitude compression function
reference channel
17
AmplitudeCompressionRefLevel,
(NR3)
Sets the amplitude compression function
target level (Vrms)
18
AmplitudeCompressionOutputLimit,
(NR3)
Sets the amplitude compression function
output limit (Vrms)
19
AmplitudeCompressionTolerance,
(NR1)
Sets the amplitude compression function
permissible error (%)
20
AmplitudeCompressionRetry, (NR1)
Sets the amplitude compression function
maximum number of retries
21
AmplitudeCompressionCorrectionFactor,
(NR1)
Sets the amplitude compression function
correction factor (%)
22
MeasIntegrationMode,
{FIX|SHORT|MED|LONG}
Sets the integration mode
23
MeasIntegrationType, {CYCLE|SEC}
Sets the integration unit to
cycles (CYCLE) or time (SEC)
24
MeasIntegrationCYCLE, (NR1)
Sets the integration period (cycles)
25
MeasIntegrationSEC, (NR3)
Sets the integration time (seconds)
26
MeasStartDelayType, {CYCLE|SEC}
Sets the measurement start delay unit
Period (CYCLE) or time (SEC)
27
MeasStartDelayCYCLE, (NR1)
Sets the measurement start delay.
(cycles)
28
MeasStartDelaySEC, (NR3)
Sets the measurement start delay time
(seconds)
29
MeasDelayType, {CYCLE|SEC}
Sets the measurement delay unit
Period (CYCLE) or time (SEC)
30
MeasDelayCYCLE, (NR1)
Sets the measurement delay period
(cycles)
31
MeasDelaySEC, (NR3)
Sets the measurement delay time
(seconds)
32
MeasSlopeCompensation, {ON|OFF}
Sets the slope compensation function to
enabled (ON) or disabled (OFF)
33
MeasCalculus, {-2|-1|0|1|2}
Sets the measurement differentiation and
integration calculations
-2: double integral
-1: integral; 0: OFF
1: differential
2: second-order differential
34
SweepLowerFrequency, (NR2)
The sweep frequency lower limit (Hz)
35
SweepUpperFrequency, (NR2)
The sweep frequency upper limit (Hz)
36
SweepResType, {LOG|LIN}
The sweep type
37
SweepRes, (NR1)
The sweep resolution
38
SweepRepeat, {ON|OFF}
Sets the repeat measurement to
enabled (ON) or disabled (OFF).
39
SweepFreqChangeMode,
The frequency change mode
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...