7.1 Error messages
7
-
2
FRA 51602
7.1 Error messages
The main errors that occur during the self-diagnosis process that is performed when the instrument is
turned on, during panel operations, and during normal measurement are described in this section, along
with their causes and how to deal with them.
If the instrument requires repair, please contact the NF Corporation or its authorized agent. When
requesting repair, please include the error code (error number) and a description of the error message that
was displayed.
An example error message is presented below. The error code is the number that follows "
Message:
" in
the information window and begins with a minus sign. In the example shown below, the error code is
"-2048". Please note that there are message codes (see "
7.2 Information messages
") that have the same
number as an error code but do not have the minus sign.
There are error messages that are not described in this manual, but which may occur in events, such as
instrument malfunction due to strong external noise.
Error messages that are not described in this manual may also appear during unusual operations such as
updating the instrument firmware. If other manuals are provided, refer to them as well.
Concerning errors that only occur during remote control operation, refer to the "Remote Control Instruction
Manual".
7.1.1 Errors that occur when the instrument power is turned on
When the power is turned on, the instrument self-diagnosis routine is executed. The error messages that
may appear when a problem occurs during that time are described in
Table 7-1
.
Table 7-1
Power-up errors
No.
Error message
Description and cause
Handling
-1024 Settings load error.
Initialize and start up.
The settings were initialized because of
a failure to load settings from the
resume memory.
Turn the power off and
on again. If the error
occurs repeatedly, there
may be a problem with
the instrument that
requires repair.
Содержание FRA51602
Страница 1: ...NF Corporation GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS...
Страница 2: ......
Страница 3: ...GAIN PHASE ANALYZER FRA51602 INSTRUCTION MANUAL OPERATIONS DA00061635 001...
Страница 24: ...1 3 Operating principles 1 10 FRA51602 Blank...
Страница 31: ...2 2 Installation 2 7 FRA51602 Figure 2 1 Rack mounting kit assembly diagram JIS...
Страница 32: ...2 2 Installation 2 8 FRA51602 Figure 2 2 Rack mounting kit assembly diagram EIA...
Страница 33: ...2 2 Installation 2 9 FRA51602 Figure 2 3 Rack mounting kit dimensions JIS...
Страница 34: ...2 2 Installation 2 10 FRA51602 Figure 2 4 Rack mounting kit dimensions EIA...
Страница 130: ...3 6 Initial settings 3 92 FRA51602 Blank...
Страница 264: ...9 9 Miscellaneous specifications 9 18 FRA51602 Figure 9 3 External dimensions...
Страница 280: ...XVI FRA51602 Blank...
Страница 282: ......
Страница 284: ......
Страница 285: ......
Страница 286: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...