Ch.3 Setup
< 94 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
Combinations of film size, type and mounted supply unit, which can be accepted
and saved are as follows;
1. Film sizes that may be set for mammography film type.
- 11x14, 10x12, 8x10
: Settings accepted
- 14x17, 14x14
: Settings not accepted
2. Relationship between supply tray types and film sizes.
- 14" supply tray
: Settings of 14X17, 14X14, 11X14
accepted.
- 10" supply tray
: Setting of 10X12 only accepted.
- 8" supply tray
: Setting of 8X10 only accepted.
The following two SCU settings must be changed when film setup changes have
been made.
- Size of film used
- Tray empty notification
p.96 SCU Setup
The message shown at the right will be
displayed if settings are incorrect.
"Data setting is incorrect."
The type specified is inappropriate for the
set film size.
"Failed to store because the value was
different from the machine setup. It will
be returned to previous value."
The film size specified is inappropriate for
the mounted supply tray. For details, refer to
the below.