ADJ USTMENTS
MODEL 3562
3-2
3-3 SAfETY CONSIDERATIONS
Although the HP 3562A is designed in accordance with international safety standards,
this manual contains information, cautions, and warnings which must be followed to ensure
safe operation and to keep the unit in a safe operating cond ition. Service and adjustments
shou ld be performed only by qualified personnel who are aware of the hazards involved.
WARNING
Any interruption of the protective (ground) conductor inside or
outside the unit or disconnection of the protective earth terminal
can expose operators to potentially dangero us voltages.
Only fuses of the rated current and specified type should be used
for replacement. The use of repaired fuses and short circu iting of
fuse holders is not permitted. Whenever fuse protection has been
impa ired, the HP 3562A must be made inoperative.
Adjustments performed in this section are performed with power
applied and the protective co vers remo ved. These adjustments
sho uld be performed only by trained service personnel who are
aware of the hazards in volved (such as fire and electrical shock).
Under no circumstances
should an operator remove any co vers,
screws, sh ields or in any other wa y access the interior of the
HP 3562A. There are no operator controls inside the instrument.
3-4 POWER SUPPLY ADJUSTMENTS
Description:
The only power supply (A1 8) adjustment is the line level voltage at which the power supply
becomes active. This adjustment should be made only when there has been a failure of
the low-line power shut down circuit or the power shut down occurs for line voltages above
81 Vrms when operated on 1 1 5 Vac mains power or above 1 62 Vrms when operated on
230 Vac mains power.
None of the H P 3562A dc power supply voltages are adjustable.
Equ ipment Required:
Variable ac power su pply
Oscilloscope
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
HP 1 980B
1 0:1 Oscilloscope probe
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
H P 1 001 4A
Содержание 3562A
Страница 2: ......
Страница 6: ......
Страница 16: ...GEN ERAL INFORMATION MODEL 3562 T bJe 1 3 Specifications cont 1 10 ...
Страница 20: ......
Страница 24: ......
Страница 126: ......
Страница 128: ......
Страница 149: ...IODEL 3562A R1 0 A22 Board Location Key Figure 3 1 7 Analog source component locator A30R 1 0 ADJUSTMENTS 3 21 3 22 ...
Страница 150: ......
Страница 152: ......
Страница 160: ......
Страница 196: ......
Страница 198: ......
Страница 206: ......
Страница 207: ...MODEL 3562A CR Cl ...
Страница 209: ...MODE L 3562A Cl ...
Страница 211: ... ...
Страница 213: ...r A1 a pQWERSuP PLY I 03562 66518 REV A REV 8 8MPOUT 58 58 FRONT REAR P ANEL ii O N ...
Страница 214: ... C401 8S1 15ISI t 1 J400 ...
Страница 217: ... ...
Страница 220: ......
Страница 221: ... ...
Страница 224: ...A3 CQVLCLI A3 ...
Страница 231: ...S V InO 3J nOS N I l3 3 1 1 1X3 NI 31dWVS lX3 H l 1 3 NNVH I 0 Ioe J ...
Страница 237: ...Sample Clock SAMP DS DATA oe _ _ _ _ _ BUS NOTE FULL SPA BASi BAI MODE ...
Страница 246: ......
Страница 259: ...WRITEL A22L l cc E Vl 8 MHz A23L ASL Inverting A1L A21 L D rivers r ...
Страница 305: ...c c I O Sequencer S r J Sequence Decoder ...
Страница 311: ... FROM 7 1 A2 SYSTEM CPU Dcf L TO 07L FROM A2 SYSTEM CPU ADDRESS L I N ES 7 1 DTACKL A2 SYSTEM CPU IRQTL ...
Страница 320: ......
Страница 321: ... From Digital Source From Digital Source Serial Data Serial Data Sinewave Interface Front End Interface 3 S D D ...
Страница 324: ...EXT TRIGGER TRIG 1 TRIG 2 CALTRIG CNTLD COMI W CNTCLK LDTRGL RESETL ...
Страница 327: ... RH r I N EXT BUFFER SAMPLE _ _ _ IN ...
Страница 425: ...L_ FAULT ISOLATION Instrument Operational Figure 7 1 3 SELF TEST Sequence MODEL 3562A STOP Enter failure in r t be 7 61 ...
Страница 450: ......
Страница 488: ......
Страница 492: ......
Страница 536: ......
Страница 552: ......
Страница 570: ......