272
────────────────────────────────────────────────────
20.2 Function Specifications
────────────────────────────────────────────────────
③ Motor power (
P
m measurement)
Measurement type
Digital computation from the measured voltage or pulse signal.
In the case that the 9603 channel A unit is torque and channel B
unit is rotation count/rate.
Accuracy
±
1dgt. with respect to the value computed from the measurement
values
Display range
0.1 to 130% of setting voltage range when both channel A and B
is set to DC voltage measurement
0 to 100% of using frequency measurement range when channel B
is set to pulse measurement.
Unit indication
W
Calculation
The calculation is carried out as follows, depending on the torque
units set for channel A.
Unit of
chA
Motor power calculations (W)
N・m
2×π×(display value of chB)
(Display value of chA)× ─────────────
60
mN・m
2×π×(display value of chB)
(Display value of chA)× ─────────────
60×100
kN・m
2×π×(display value of chB)×1000
(Display value of chA)× ────────────────
60
kgf・m
2×π×(display value of chB)×9.80665
(Display value of chA)× ────────────────
60
kgf・cm
2×π×(display value of chB)×9.80665
(Display value of chA)× ────────────────
60×100
7. Frequency measurement (
fa, fb, fc)
(when using the 9600, 9601, 9602)
Measurement type
Reciprocal calculation from frequency of input waveform
Number of measurement
channel
3 channels
Display range
AUTO/ 50 Hz/ 500 Hz/ 5 kHz/ 50 kHz/ 2 MHz
Resolution
99999 counts
Measurement range
500.00 mHz to 50.000 Hz/ 20.000 Hz to 500.00 Hz/
200.00 Hz to 5.0000 kHz/ 2.0000 kHz to 50.000 kHz/
20.000 kHz to 2.0000 MHz
Measurement accuracy
±
0.1% rdg.
±
1dgt.
With a sine wave input from 10% to 130% of the
U
/
I
ranges,
when the frequency characteristics of the input unit forming the
source are within the specified range
Function
Switching
U
/
I
source
(Simultaneous
U
/
I
input from a single input unit is not possible.)
Filter switching linked to range (high-pass or low-pass filter)
Pulse measurement when 9603 is used
When the 9303 is set to pulse measurement, fc is forced to pulse
measurement of channel B.
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......