16.4 Internal Block Diagram
────────────────────────────────────────────────────
U
±
±
Attenuation
circuit
Buffer
Range
Range
Isolation
amplifier
HPF
LPF
RMS/DC
MEAN/DC
A/D
3193
3193
-10
Switch control
RMS/DC
MEAN/DC
Multiplication
/DC
HPF
LPF
Auto-zero
Peak hold
Waveform
shaping
Polarity
decision
Over
detect
f measurement
U
/
I
D/A
LEAD/LAG
OVER
I
DMAG (0 Adj)
Sensor
OVER
U
16.4 Internal Block Diagram
The voltage value is converted by attenuator and range circuits to a voltage
signal waveform proportional to the measured voltage, then isolated by an
isolating amplifier.
The sensor input is converted in a range circuit to a voltage signal waveform
proportional to the measurement current.
For details on RMS value, MEAN value, active power, waveform peak value
measurement circuit, peakover detect circuit, see Chapter 14.
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......