251
────────────────────────────────────────────────────
16.3 Specifications (using with the HIOKI 3193)
────────────────────────────────────────────────────
16
17
18
19
5
6
7
8
9
10
11
12
13
14
A
Voltage (
U
)
Current (
I
)
Active power (
P
)
Influence of maximum
rated voltage to earth
(600 Vrms at 50/60 Hz)
±
0.05%f.s. max
(When applying between voltage input terminal and case)
Dielectric strength
(50/60 Hz)
5.55 kV AC
(
1 mA
)
for 1 minute (When using with the 3193)
Between
U
input terminals and case, between
U
input terminals and power
plug, between
U
input terminals and sensor input terminal
Insulation resistance
100 M
Ω
or more at 500 V DC (When using with the 3193)
Between
U
input terminals and sensor input terminal, between
U
input
terminals and case, between
U
input terminals and power plug
Input
voltage
(
U
)
Frequency (Hz)
9602 Voltage Accuracy Assurance Range
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......