78
────────────────────────────────────────────────────
7.8 Zero suppress function
────────────────────────────────────────────────────
F1
F2
INTEG-ZERO
F3
NOTE
7.8 Zero suppress function
You can set the minimum value of data to be integrated. You can change
the value if the input level is too low for the preset range.
1. Press the
STATUS
key, then use the
PAGE
(
)
keys to display the "
SYSTEM
" page.
2. Using the
CURSOR
keys, move the cursor to the
"
INTEG-ZERO
" item.
3. Select from
F1
(0.5%)
,
F2
(0.1%)
,
F3
(0.0%)
.
4. Press the
MEAS
key to return to the measurement
screen.
・
In zero suppress, figures less than the percentage set for the full range are
regarded as "0".
・
The setting is 0.5% by default.
・
When the setting is at 0.0% or 0.1%, the values not displayed as
instantaneous values will also be counted. Counts may be displayed even
when there is no input due to internal off-set.
・
This change function is not transferred when data is saved or loaded via PC
communications or floppy disk. Therefore, settings must be made manually
before measurement.
・
The zero suppress in the display of instantaneous values is not changed.
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......