4.10 Setting the Averaging
────────────────────────────────────────────────────
10-3.TIF #CHANGE
F1
F2
F3
F4
AVE TIME
TIME
NOTE
1. Press the
STATUS
key, then use the
PAGE
(
)
keys to display the "
TIME
" (time control) page.
2. Using the
CURSOR
keys, move the cursor to the
"
AVERAGING
" item.
3. Press
F2
(
ave time
).
4. Select desired time setting from interval time, timer
time, real control time. See Section 7.2.
5. Press the
MEAS
key to return to the measurement
screen. Pressing the
START/STOP
key starts
averaging.
6. To return to the normal measurement, release the
hold state, because the operation stops in hold state
("HOLD" indicates) when the time averaging is
completed. When used together with the interval
time, the hold value is the average within the final
interval time. To switch display to the total
average, press the
HOLD
key. The TOTAL mark
appears during display of the total average.
7. To complete forcibly, press the
START/STOP
key
again.
・
The time average requires a setting of an interval control time, timer control
time, or real-time control time.
・
Operation with the START/STOP key also affects integration. Therefore,
after completion of averaging or following forced termination, press the
SHIFT key, then press the START/STOP key to reset integration.
・
When used in conjunction with a timer control time or real-time control
time, when the set time elapses the overall average value is displayed and
held, which the unit stops. When this hold state is ended, the unit returns to
normal measurement.
・
For the time average function, floppy disk and printer settings are also
synchronized in the same way.
・
During the time average function operation, it is not possible to change
settings.
・
If an interval control time is set and the
HOLD
key is pressed, the display is
updated each time the interval elapses.
・
During the averaging operation, if an out-of-range data value is included, the
unit indication is red.
・
No polarity is shown for time averaging when [TYPE1] is selected as the
calculation type for reactive power (Q).
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......