68
────────────────────────────────────────────────────
7.1 Overview
────────────────────────────────────────────────────
NOTE
・
Data for each interval of the interval time setting is displayed on the screen
in hold state. When the value is not held, it must be combined with the
floppy disk drive or printer function to display.
・
Calculation results (DC voltage) from the various input units are integrated
at the rate of 64 samples per second. Therefore, integration results may vary
with instruments that use different response speeds, sampling rates or
calculation methods.
・
During integration operation, if the integration item is "o.r" (out-of-range),
the value of 130% of range is integrated. In this case, the integration value
indication is red.
・
In the DC mode, current is integrated from instantaneous current waveforms.
In the AC+DC/AC mode, it is integrated as RMS or MEAN values.
・
In the DC mode, power is integrated from instantaneous power waveforms.
In the AC+DC/AC mode, it is integrated as effective power.
・
Once integration starts, items for which auto ranging is set have their ranges
fixed from the start time. Set the range so that it will not be exceeded before
beginning integration.
・
During integration (even "on standby" under real-time control), all settings
except changing screens and the hold and peak hold functions are disabled.
・
While the hold function is activated, the display is frozen, but internally the
integration continues normally. When the floppy disk drive or D/A outputs
are used, however, the held values continue to be output.
・
While the peak hold function is activated, the integration operation is not
affected.
・
After integration ends, until the integration values are reset, it is not possible
to change settings.
・
The calculation results (for DC voltage) from each input units are integrated
64 times per second.
Operation Procedure
1. Set the required time settings (interval, timer, or real-time control).
For the manual integration, make the time settings to OFF, however, the
timer time operates as 10000 hours.
2. Set the external output for FDD, printer, D/A if necessary.
See Chapter 10, Chapter 11, Chapter 13
3. Start integration.
Press the
START/STOP
key.
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......