38
────────────────────────────────────────────────────
4.3 Switching the Voltage Range and Current Range
────────────────────────────────────────────────────
6 V ⇔ 15 V ⇔ 30 V ⇔
60 V ⇔ 150 V ⇔ 300 V ⇔
600 V ⇔ 1 kV
RANGE
U
I
F1
F2
F5
UNIT
U RANGE
NOTE
Switching the range on each channel screen.
1. Switch to display the channel for which you wish
to change the setting.
2. Hold down the panel
RANGE
(+,-) key until the
desired setting range is displayed.
3. To set auto ranging, hold down the panel
RANGE
(+) key or press both (+,-) keys simultaneously.
4. To cancel auto ranging, press either of the
+
and
-
keys.
Switching on the STATUS screen
1. Press the
STATUS
key, then use the
PAGE
(
)
keys to display the "
UNIT
" page.
2. Using the
CURSOR
keys, move the cursor to the
channel to be changed of the "
U RANGE
" or "
I
RANGE
" item.
3. Select range from
F5
(AUTO)
,
F1
(
↑
; range up),
F2
(
↓
; range down).
4. Press the
MEAS
key to return to the measurement
screen.
・
For channel combinations of 1P3W and above, the channels are forced to the
same range. In this case the range for the lowest-numbered channel is used
for all of the channels.
・
The auto ranging function switches up a range when a measurement value
exceeds 110% of measurement range (out-of-range) or when a waveform
peak exceeds six times the range value (peak over), and switches down a
range when the value is less than 30% of the nominal range. When
measuring a distorted waveform the range selection may not be stable. In
this case use manual range setting.
・
When the integration function time average has started, auto ranging is
disabled, and the range remains fixed from that point.
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......