167
────────────────────────────────────────────────────
:INTEGrate:RESEt
────────────────────────────────────────────────────
Syntax
:INTEGrate:RESEt
Example
:INTEGRATE:RESET
Resets integration value for all
channels.
Syntax
:INTEGrate:STARt
<channel no.>
<channel no.> = 1 to 6
Example
:INTEGRATE:START
Starts simultaneously integration for all
channels.
:INTEGRATE:START 1
:INTEGRATE:START 2
:INTEGRATE:START 3
The integration start timing is offset,
and the start is for channels 1, 2, and
3 in that order.
Syntax
:INTEGrate:STOP
<channel no.>
<channel no.> = 1 to 6
Example
:INTEGRATE:STOP
Stops simultaneously integration for all
channels.
:INTEGRATE:STOP 1
:INTEGRATE:STOP 2
:INTEGRATE:STOP 3
The integration stop timing is offset,
and the start is for channels 1, 2, and
3 in that order.
Function
Resets the integration value.
Note
The integration values for all channels are
simultaneously reset. In this case, for the
channel during integration, an execution
error occurs.
Function
Starts integration for the specified
channel or all channels.
When the channels are specified, the
integration for the specified channel
starts.
When the channel is not specified, the
integration for all channels starts.
Notes
・
The items to be integrated are the
specified items by the :DISPlay:
INTEGrate [channel no.] <A> command.
・
For the detail screen for is 1P3W or
above, for the channel specification enter
the number of the lowest-numbered
channel in the combination. For example,
when using input unit channels 1 to 3 in
3P4W mode, specify "1".
・
When the interval time, timer time, or
real time control time is set and if the
integration starts with all channels, the
operation is controlled by the specified
time, however, if the channels are
specified, time controls are invalid to
operate manually.
Function
Stops integration for the specified
channel or all channels.
When the channels are specified, the
integration for the specified channel
stops.
When the channel is not specified, the
integration for all channels stops.
Note
If the channel is not specified to stop, the
integrations for all channels are
simultaneously stopped even when the
integration started with specified channel,
:INTEGrate:RESEt
Resets the integration value.
:INTEGrate:STARt
Starts the integration.
:INTEGrate:STOP
Stops integration.
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......