141
────────────────────────────────────────────────────
:AOUT
────────────────────────────────────────────────────
Syntax
AOUT
<A, ...8 items>
<A> =
U
n
, I
n
, P
n
, Q
n
, S
n
, PF
n
,
DEG
n
, PIH
n
, MIH
n
, IH
n
,
PWP
n
, MWP
n
, WP
n
, LF
n
,
FA, FB, FC, EFF1, EFF2,
EFF3, EXTA, EXTB, PM,
PK1, PK2, PK3, PK4, PK5,
PK6
n
: 1 - 6 (channel 1 to 6),
12 (channels 1 and 2),
34 (channels 3 and 4),
45 (channels 4 and 5),
56 (channels 5 and 6),
123 (channels 1, 2, and 3),
456 (channels 4, 5, and 6)
(depends on combination of
channels)
Example
:AOUT PF1,S2,FA
D/A output channel 1 is assigned to
PF1 (power factor calculated from
input unit channel 1), D/A output
channel 2 is assigned to S2 (apparent
power calculated from input unit
channel 2), and D/A output channel 3
is assigned to FA (frequency
calculated from input unit channel 1).
Syntax
AOUT?
Response
syntax
Same as the :AOUT command.
AOUT
<A, ...(8 items)>
Example
Transmission
Response
AOUT?
AOUT PF1,S2,FA,U1,U1,U1,U1,U1
Function
Sets the output items for channels 1 to 8
respectively of the D/A outputs
The D/A output item which is not
specified is set to
U
1.
Errors
・
Execution error / If the setting data is an
item which cannot be selected because of
the number of input units.
・
Command error / If the setting data is not
character data listed on the left.
Note
U: Voltage
I: Current
P: Active power
Q: Reactive power
S: Apparent power
PF: Power factor
DEG: Phase angle
IH: Integration current
(PIH: positive, MIH:negative)
WP: Integration power
(PWP: positive, MWP: negative)
LF: Load factor
F: Frequency
EFF: Efficiency
EXT: External-In
PM: Motor power
PK: Peak value
Function
Returns the D/A output items as
character data.
Note
With this query, if any errors occurs, no
response message is produced.
The D/A output item which is not
specified is set to
U
1.
12.4.2 Specific Commands
:AOUT
Sets the items of D/A output.
AOUT?
Queries which the items of D/A output is to be performed
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......