69
────────────────────────────────────────────────────
7.2 Setting the Control Time
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
NOTE
7.2.1 Setting the Interval Time
F1
F2
F1
F2
TIME
INTERVAL
TIME
7.2 Setting the Control Time
Using the three time control functions provided by this unit, it is possible to
control time averaging, the floppy disk drive, printer, and integration
functions.
・
It is not possible to make separate settings for time averaging, floppy disk
drive, printer, and integration functions.
・
The integration function always operates. Therefore, even when the
integration value is not being measured, during the operation period of time
control, the "
INTEG
" indication appears. After end of time control, reset
(
SHIFT
→
START/STOP
) to turn off the "
INTEG
" display.
・
When using time averaging integration, some settings are disabled. If auto
ranging is set, the range is fixed at the start time. The HOLD function,
however, can still be used.
・
Even when time settings are made, the unit does not operate until the panel
START/STOP
key is pressed.
・
The operation starts by real-time control and it is completed by timer
control. In this case, the stop time by real-time control is ignored.
Interval control operates the unit repeatedly at the specified interval.
1. Press the
STATUS
key, then use the
PAGE
(
)
keys to display the "
TIME
" (time control) page.
2. Using the
CURSOR
keys, move the cursor to the
"
INTERVAL
" and press
F2
to set to ON.
3. Move the cursor to desired time digit, and set the
time using the
F1
(↑)
and
F2
(
↓
).
4. After setting completed, press the
MEAS
key to
return to the measurement screen.
5. Press the
START/STOP
key to start operation.
411-1.tif 411-2.tif CHANGE
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......