4.3 Switching the Voltage Range and Current Range
────────────────────────────────────────────────────
Current range
43-1.tif
43-2.tif
43-3.tif
Voltage range
4.3 Switching the Voltage Range and Current Range
When the voltage range and current range is displayed on the screen for each
channel, it is also possible to change the ranges directly with the panel keys.
This is also possible from the STATUS screen in the "
UNIT
" display.
Effective input range
The effective input range is 5% to 110% of range.
(for the 9600 and 9601, 5% to 100% for 1 kV range only, for the 9602, 5%
to 100% for 600 V range only)
Display range
The value which can be displayed is up to 130% of range. If the value
exceeds this range, "o.r" is displayed.