35
────────────────────────────────────────────────────
4.1 Setting the Wiring Mode (1P2W to 3P4W)
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
NOTE
Channels
1
2
3
4
5
6
①
1P2W
1P2W
1P2W
1P2W
1P2W
1P2W
②
1P3W / 3P3W
1P2W
1P2W
1P2W
1P2W
③
1P3W / 3P3W
1P3W / 3P3W
1P2W
1P2W
④
1P3W / 3P3W
1P3W / 3P3W
1P3W / 3P3W
⑤
3V3A / 3P4W
1P2W
1P2W
1P2W
⑥
3V3A / 3P4W
1P3W / 3P3W
1P2W
⑦
3V3A / 3P4W
3V3A / 3P4W
Measurement line
Mode
Display item
Single-phase two-wire
(1φ2W)
1P2W
U, I, P, Q, S, λ/φ, |Up|/|Ip|
Single-phase three-wire
(1φ3W)
1P3W
(channels 1+2)
U1, U2, U12, I1, I2, I12, P1, P2, P12, Q1, Q2, Q12, S1, S2,
S12, λ1/φ1, λ2/φ2, λ12/φ12, |U1p|/|I1p|, |U2p|/|I2p|
Three-phase three-wire
(3φ3W)
3P3W
(2 voltages, 2 currents,
2 power meters method)
(channels 1+2)
U1, U2, U12, I1, I2, I12, P12, Q12, S12, λ12/φ12,
|U1p|/|I1p|, |U2p|/|I2p|
3V3A
(3 voltages, 3 currents,
2 power meters method)
(channels 1+2+3)
U1, U2, U3, U123, I1, I2, I3, I123, P123, Q123, S123,
λ123/φ123, |U1p|/|I1p|, |U2p|/|I2p|, |U3p|/|I3p|
Three-phase four-wire
(3φ4W)
3P4W
(channels 1+2+3)
U1, U2, U3, U123, I1, I2, I3, I123, P1, P2, P3, P123, Q1, Q2,
Q3, Q123, S1, S2, S3, S123, λ1/φ1, λ2/φ2, λ3/φ3,
λ123/φ123, |U1p|/|I1p|, |U2p|/|I2p|, |U3p|/|I3p|
・
It is only possible to select from the combinations shown in the connection
setting screen.
For combinations 1P3W and above, adjacent units must be of the same type.
・
When using the 9602
AC/DC CLAMP INPUT UNIT
, only a combination of the
same current sensor type can be selected. In other cases, all are set to 1P2W.
・
When using the 9602
AC/DC CLAMP INPUT UNIT
, if the sensor
configuration is changed (including with the sensor not connected), and the
unit is then powered on, "Resetting due to configuration change." is
displayed. Pressing
F1
(YES)
resets settings. Pressing
F2
(NO)
does not
reset and settings remain unchanged, but the display value for current may
be changed. Return the configuration settings and power on again. The
message is not displayed.
・
When using clamp-on units, if the sensor rating is changed and the unit is
then powered on, in the 1P2W mode the CT ratio within the unit is
automatically set accordingly. For combinations 1P3W and above, if other
channels in the combination have different ratings the combination is
disabled for measurement.
・
The display items in the various modes are as follows.
・
Although the display will show for each channel
3P3W
or
3V3A
the active
power (
P1, P2 and P3
), reactive power (
Q1, Q2 and Q3
), apparent power
(
S1, S2 and S3
), power factor (
λ
1,
λ
2 and
λ
3
), and phase angle (
φ
1,
φ
2
and
φ
3
), please be aware that these figures have no meaning.
・
Power factor (
λ
) and phase angle (
φ
) are not be displayed simultaneously.
・
The measurement values for three-phase three-wire in 3P3W and 3V3A mode
are same because of same measurement method.
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......