4.10 Setting the Averaging
────────────────────────────────────────────────────
Z
(n-N-1)
+ Z
(n-N-2)
+...+ Zn
Display value =
――――――――――――――
N
Zn: nth measured data
N: number of setting samples
MOVING AVE
TIME
F1
F2
F3
F4
TIME
NOTE
(2) Setting the Moving Average (AV-M)
The moving average function displays a simple average calculated by
summing the measurement values from the beginning of averaging, and
dividing by the number of samples, until the specified number of samples.
From that point on it discards the oldest data value as each new value is
added, thus yielding a simple average over the most recent specified number
of samples.
1. Press the
STATUS
key, then use the
PAGE
(
)
keys to display the "
TIME
" (time control) page.
2. Using the
CURSOR
keys, move the cursor to the
"
AVERAGING
" item.
3. Press
F3
(
Moving ave
).
4. Move the cursor to the specified number of
sampling item on the right, and set the number of
sampling using the function key.
5. Press the
MEAS
key to return to the measurement
screen, and averaging starts.
This has no connection with any time settings.
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......