191
────────────────────────────────────────────────────
:TIMER?
────────────────────────────────────────────────────
Syntax
:TIMER?
Response
syntax
:TIMER:CONTROL
<ON/OFF>
;
TIME
<hour,min>
Example
Transmission
Response
:TIMER?
:TIMER:CONTROL ON;TIME 00000,10
Syntax
TRANsmit:COLumn
<NR1>
<NR1> = 0, 1
Example
Transmission
Response
Transmission
Response
When :TRANsmit:COLumn 0
:MEAS? U1,I1
78.01E+00;5.012E+00
When :TRANsmit:COLumn 0
:MEAS? U1,I1
+078.01E+00;+05.012E+00
Syntax
TRANsmit:COLumn?
Response
syntax
Headers: ON
:TRANSMIT:COLUMN
<0/1>
Headers: OFF
<0/1>
Example
Transmission
Response
(headers on)
(headers off)
:TRAN:COL?
:TRANSMIT:COLUMN 0
0
Function
Queries the current settings for timer
control.
Function
Sets the numerical data format of
:MEASure?
If <NR1> = 0, the leading zero is omitted
from the mantissa of NR3 numeric data.
If <NR1> = 1, the number of NR3
numeric data is constant (the leading zero
is not omitted from the mantissa).
Note
・
<NR1> can be accepted in NRf format,
but decimal fractions are rounded off.
・
This has no effect on integration time
(TIME), display blanking, calculation
impossibility, out of range values.
・
The setting of powering on is 0.
Error
Execution error/ If data is set other than
NRf format.
Function
Queries the current setting of numerical
data format of :MEASure?.
The returned numerical value corresponds
to the setting state of the NR3 numeric
data as follows:
If <NR1> = 0, the leading zero is omitted
from the mantissa of NR3 numeric data.
If <NR1> = 1, the number of NR3
numeric data is constant (the leading zero
is not omitted from the mantissa).
Note
With this query, if any error occurs, no
response message is produced.
Errors
・
Query error/ If the response message is
longer than 2000 bytes.
・
System error/ This query is not executed.
A device dependent error occurs.
:TIMER?
Queries the timer control.
:TRANsmit:COLumn
Select the numerical data format.
:TRANsmit:COLumn ?
Queries the numerical data format.
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......