14.2 Notes on Use
────────────────────────────────────────────────────
14.2 Notes on Use
・
The 9600 is a factory-fitted option. It therefore requires the 3193 unit for
calibration or repair.
・
For accurate measurement, allow one hour for warming up before use.
・
It may not be possible to obtain accurate measurements close to a
transformer or conductor carrying a large high-frequency current, or close to
any device such as a radio transmitter generating a strong magnetic field.
・
The 9600 uses the DC-CT method for measuring current, and therefore after
measuring a large current, there may be a very small residual offset voltage.
The offset voltage produces the largest error effect in the minimum ranges;
in this case, shut off the current input, and carry out degaussing (DMAG).
・
The 9600 active power measurement unit operates with an auto-zero circuit
at 2.442 kHz. For this reason, an input signal with a frequency of 2.442 kHz
will result in a periodically fluctuating display indication.
・
When measuring a high frequency voltage to earth (for example the
secondary side of an inverter), errors may occur in the measurement values.
・
Limits are specified for the range in which voltage and current level
accuracies are guaranteed, depending on the input frequency.
・
The guaranteed accuracy ranges of frequency depend on the response,
coupling mode, and low-pass filter settings.
・
Depending on the response setting, display may not stabilize during
measurement of low frequencies. If this occurs, use averaging.
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......