10.3 Output Rate
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Selecting items for output
Full scale
Voltage value, current value for each channel,
SUM value of voltage and current,
9603 External signal unit
(U1 - U6, I1 - I6, U12, U34, U56, U123, U456, I12,
I34, I56, I123, I456, Va, Vb)
Measurement range
Effective power, reactive power, apparent power for
each channel
(P1 - P6, Q1 - Q6, S1 - S6)
(voltage range)×(current range)
For example, when measuring in the 300 V range and 10
A range, 3 kW corresponds to full scale.
Display value: -3 kW - 0 - +3 kW
D/A output value: -5 V - 0 - +5 V
Effective power, reactive power, SUM value of
apparent power during measuring 1P3W, 3P3W,
(P12, P34, P56, Q12, Q34, Q56, S12, S34, S56)
Effective power, reactive power, SUM value of
apparent power during measuring 3V3A
(P123, P456, Q123, Q456, S123, S456)
(voltage range)×(current range)×2
For example, when measuring in the 300 V range and 10
A range, 6 kW corresponds to full scale.
Display value: -6 kW - 0 - +6 kW
D/A output value: -5 V - 0 - +5 V
Effective power, reactive power, SUM value of
apparent power during measuring 3P4W
(P123, P456, Q123, Q456, S123, S456)
(voltage range)×(current range)×3
For example, when measuring in the 300 V range and 10
A range, 9 kW corresponds to full scale.
Display value: -9 kW - 0 - +9 kW
D/A output value: -5 V - 0 - +5 V
Power factor (λ)
Display value: -1 - 0 - +1
D/A output value: -5 V - 0 - +5 V
Phase angle (φ)
Display value: -180
。
- 0 - +180
。
D/A output value: -5 V - 0 - +5 V
Efficiency (η)
Display value: 0 - 100%
D/A output value: 0 - 5 V
Current integrate value (Ih)
(current range)×(integrated time)
For example, when integrating in the 10 A range for 1
hours, 10 Ah corresponds to full scale.
Display value: -10 Ah - 0 - +10 Ah
D/A output value: -5 V - 0 - +5 V
Effective power integrate for 1P2W (WP)
(voltage range)×(current range)×(integrated time)
For example, when integrating in the 300 V range and 10
A range for 1 hours, 3 kWh corresponds to full scale.
Display value: -3 kW - 0 - +3 kW
D/A output value: -5 V - 0 - +5 V
Effective power integrate for 1P3W, 3P3W, 3V3A
(WP)
(voltage range)×(current range)×(integrated time)×2
For example, when integrating in the 300 V range and 10
A range for 1 hours, 6 kWh corresponds to full scale.
Display value: -6 kW - 0 - +6 kW
D/A output value: -5 V - 0 - +5 V
Effective power integrate for 3P4W (WP)
(voltage range)×(current range)×(integrated time)×3
For example, when integrating in the 300 V range and 10
A range for 1 hours, 9 kWh corresponds to full scale.
Display value: -9 kW - 0 - +9 kW
D/A output value: -5 V - 0 - +5 V
Frequency (f)
Frequency range is full scale.
NOTE
10.3 Output Rate
The D/A outputs are
±
5 V DC corresponding to full scale, where the full
scale values are as shown in the following table.
・
For integration, the integration time is the time interval set for the timer time
or the real-time control time.
・
For manual integration, the integration time is output as 10,000 hours.
Содержание Power HiTester 3193
Страница 1: ...3193 3193 10 Instruction Manual POWER HiTESTER EN Feb 2019 Revised edition 16 3193A981 16 19 02H ...
Страница 2: ......
Страница 50: ...32 3 9 Operations During Power Failure ...
Страница 76: ...58 4 13 Degaussing ...
Страница 80: ...62 5 2 Setting the Frequency Range fa ...
Страница 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Страница 112: ...94 10 3 Output Rate ...
Страница 250: ...232 13 9 Error and Overflow Displays ...
Страница 278: ...260 17 5 Internal Block Diagram ...
Страница 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Страница 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Страница 306: ...INDEX 4 Index ...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......