Section 7-6 - Minimum Cable Configuration to Boot
Section 7-6
Minimum Cable Configuration to Boot
Following Cables must be connected for minimum Configuration:
see:
Figure 7-11: Cable- Minimum configuration
a.) RS232-1 (connector for User Interface, near to the Mouse connector)
NOTE:
Don´t mix up the RS232-1 with the RS232-2 and RS232-3 connectors.
b.) VGA (Monitor)
c.) CPN (Primary Power Supply)
d.) Console
e.) Stand By switch
f.) Monitor Power Supply
g.) Mouse (connector for Mouse/Keyboard)
Figure 7-11 Cable- Minimum configuration
g
a
b
c
d
e
f