Section 7-7 - Troubleshooting Trees and Instructions
Section 7-7
Troubleshooting Trees and Instructions
7-7-1
System does not Power On / Boot Up
Figure 7-12 System does not boot up
Is the scanner pluged
in?
Is there AC at the wall
outlet?
Make sure the
Scanner isn't
plugged into
emergency power
(red outlet)
Yes
Is the circuit breaker on
Yes
+ These three condidtions should be the first items checked if
the scanner doesn't boot up.
+ If all three conditions are present, the scanner should be in
astandby mode with the ON/OFF switch illuminated with amber
color.
+ Follow the flow chart to determine the reason the scanner
doesn't boot up
No
No
Switch in ON
Position
Fan Noise ?
Yes
+ Check if the power switch over the mains plug of the Primary Power
Supply isn't switched off.
+ Check the connections to and from the Power Supply
+ Check the power of the cable which connects the Power Supply with
the GEF module (description Check Points Voltages)
+ If ther is no power and the connections are O.K., replace the Primary
Power Supply
No
Yes
System is on and ready for use
Section 7-3
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......