GE M
EDICAL
S
YSTEMS
- K
RETZTECHNIK
U
LTRASOUND
D
IRECTION
105844, R
EVISION
1
V
OLUSON
® 730 S
ERVICE
M
ANUAL
3-2
Section 3-1 - Overview
3-1-3
Installation Warnings
1.) Since the Voluson® 730 weighs approximately 136 kg (300 lb.) without options, preferably two
people should unpack it. Two people are also preferable for installing any additional items in excess
of 35 pounds (e.g. Monitor).
2.) There are no operator serviceable components. To prevent shock, do not remove any covers or
panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service
personnel should carry out servicing and troubleshooting.
3.) After being transported, the unit may be very cold or hot. If this is the case, allow the unit to
acclimate before you turn it on. It requires one hour for each 2.5
°
C increment it's temperature is
below 10
°
C or above 40
°
C.
3-1-3-1
Brake Pedal Operation
3-1-3-2
Operator I/O Panel Position
If weight is placed on the Operator I/O Panel in it’s extended position the console could tip over.
CAUTION
Equipment damage possibility. Turning the system on without acclimation after arriving at site
may cause the system to be damaged.
Table 3-3
Acclimation Time
°
C
60
55
50
45
40
35
30
25
20
15
10
5
0
-5
-10
-15
-20
-25
-30
-35
-40
°
F
140
131
122
113
104
96
86
77
68
59
50
41
32
23
14
5
-4
-13
-22
-31
-40
hrs
8
6
4
2
0
0
0
0
0
0
0
2
4
6
8
10
12
14
16
18
20
Figure 3-1 pulling, moving or lifting the system
WARNING
WARNING
REMEMBER: If the front wheels are engaged for transportation, pressing the release
brake pedals (brakes on front wheels under the foot rest) once disengages the lock.
WARNING
WARNING
The system should NOT be moved with the Operator I/O Panel extended. Move the
Operator I/O Panel to it’s centered and locked position.
WARNING
WARNING
Monitor mounting mechanism may break if not properly supported (e.g. with packing
foam) during transportation even inside the hospital.
When pulling, moving or lifting the system,
grasp it only at the rear handle of the trolley and
the handle underneath the foot rest.
Do
NOT
pull or lift the system with the front handle
of the user interface (operator panel).
WARNING
WARNING
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......