Chapter 4 Functional Checks
4-3
4.) Press the
ON/OFF
Standby switch left below the Control Panel once.
When the
ON/OFF
Standby switch left below the Control Panel is pressed once, the System (including
the Back-end Processor) starts and the software code is distributed to initiate the scanner.
The company logo appears on the Touch Panel during boot up.
Depending on the BIOS-Version no status messages are displayed during this process.
Boot up time is about 3 minutes.
NOTE:
The mains outlet of the system for peripheral auxiliary equipment are commonly switched with
the Standby switch. So the auxiliary equipment need not to be switched ON/OFF separately.
4-3-2
Power Off / Shutdown
NOTE:
After turning off a system, wait at least ten seconds before turning it on again. The system may not be
able to boot if power is recycled too quickly.
4-3-2-1
Scanner Shutdown
1.) Press the
ON/OFF
Standby switch left below the Control Panel once.
2.) Switch OFF the Circuit Breaker at the rear of the system.
3.) Disconnect the Mains Power Cable is necessary.
For example:
Relocating the scanner.
Figure 4-2 ON/OFF Stand-By Switch
Figure 4-3 Circuit Breaker and Power Cable on Back of Voluson® 730
ON/OFF Standby Switch
Circuit Breaker
Mains Power Cable
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......