3-6
Section 3-3 - Preparing for Installation
Section 3-3
Preparing for Installation
3-3-1
Connection of Auxiliary Devices
3-3-1-1
Monitor Connection
1.) Connect all the Monitor cables a shown in
Figure 3-4
.
2.) Mount connector protection at the monitor cables.
3.) Activate the power switch on the rear side of the monitor.
NOTE:
The switch of the monitor has to be in ON position before starting the system.
Leave monitor power switch always ON.
3-3-1-2
Connection of Peripherals
1.) Remove rear cover plate. see
Figure 3-5
-
Carefully remove plastic caps using a slotted screwdriver or the blade of a knife.
-
Loosen screws and washers.
2.) When the cover is loose on top, pull the rear cover out and move upwards.
Figure 3-4 Monitor Connection
Figure 3-5 Remove rear cover plate
Connector protection
of monitor
Power switch
plastic caps; screws and washers