Chapter 3 Installation
3-27
3-5-5-3
CP770DW - Printer Settings
(cont’d)
7.) For saving the adjusted printer settings click OK.
NOTE:
When you call up the “Gamma Adjust” page again, the Input and Output parameters won´t be
shown until re-adjustment of the displayed Gamma curve.
8.) For saving the adjusted printer settings click APPLY and then OK.
9.) Finally close the ‘Printers’-window with the close button and exit System Setup with SAVE&EXIT.
3-5-6
Printer Remote Control Selection
To assign the Remote
PRINT A
key, Remote
PRINT B
key and Report Printer to desired Printer:
1.) On the Touch Panel, touch UTILITIES.
2.) In the Utilities menu, touch SYSTEM SETUP to invoke the setup desktop on the screen.
3.) Select the
PERIPHERALS
page.
•
Remote Print A:
Select the desired Printer for the remote control
PRINT A
key.
•
Remote Print B:
Select the desired Printer for the remote control
PRINT B
key.
NOTE:
Optionally the Remote Control can be done by Foot switches.
Therefore select “Print A” or “Print B” in “Foot Switch Left” or “Foot Switch Right” - section.
•
Report Printer:
Select the desired Report Printer from the drop-down menu.
Figure 3-36 Peripherals page
NOTICE
The selected Report Printer (usually HP 990cxi or HP 995c) is used for printing reports and images from
Sonoview.
Foot Switch
Remote
Control
Report Printer
Print B
Print A
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......