Section 4-4 - Functional Checks
4-4-5
Volume Mode Checks
(cont’d)
4-4-5-3
Functions in the Tool Menu
For further details refer to the Voluson® 730 Basic User Manual:
•
Chapter 11.2, 3D Acquisition
•
Chapter 11.3, 3D Image Rendering, Operation
•
Chapter 11.5, Real Time 4D Acquisition
•
Chapter 11.6, Live 3D Acquisition
•
Chapter 11.7, Real Time 4D Biopsy
•
Chapter 11.8, Volume Contrast Imaging (VCI)
Figure 4-13 Tool Menu
Table 4-9
Functions in the Tool Menu
Step
Task
Expected Results
1
RENDER MODE
To select the Render Mode (Image Type and Render Algorithm)
2
MAGI CUT
Ability to electronically manipulate the images and cut way “3D artifacts”.
3
GRAY CHROMA MAP
Depending on individual requirements a “harder” or “softer” image can be obtained.
4
RENDER DIRECTION
To select the desired Render Direction. The green line symbolizes the direction of the view.
5
TH. LOW
By changing the treshold all echoes below the level are enhanced in color to remove noise
and to get a “clear sight”. (Surface Mode only)
6
TRANSPARENCY
A higher number makes the gray scale information more transparent.
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......