Chapter 4 Functional Checks
4-11
4-4-3
Spectral Doppler Mode Checks
For further details refer to the Voluson® 730 Basic User Manual, Chapter 7, Spectral Doppler Mode.
Figure 4-9 PW Main- and PW Sub Menu
Table 4-5
Spectral Doppler Mode (PW, CW) Functions
Step
Task
Expected Results
1
Gate Position and Gate Size
Adjust the Gate- Position resp. Size with the
TRACKBALL
in the B-Single image.
The
upper trackball key
changes from Gate position to Gate size.
2
Activation of Doppler-Mode
Press the
right trackball key
to activate the motion display.
Press the
left trackball key
to activate both Modes (B/D).
3
Doppler-Gain
Rotate the
PW-MODE
key to adjust the amplification of the entire spectrum.
4
STEERING
The steering function is only available with linear probes.
5
SPEED
By touching up or down, four different sweep speeds can be selected.
6
RT TRACE (Real time Auto-Trace)
The envelope curve of the Doppler spectrum (maximum velocities) and the
corresponding evaluations are automatically displayed on the monitor.
7
INVERT
To invert the Doppler spectrum display in relation to the direction of the flow.
8
ANGLE
The angle cursor can be turned in both directions without stop. By pressing the
angle knob repeatedly the angle correction switches from +60° to 0° and to -60°.
9
BASELINE
Adjusting the baseline is possible in Read- and Write-Mode (up/down in 8 steps).
10
WMF
(Wall Motion Filter)
Used to eliminate Doppler “noise” that is caused by vessel wall motion.
11
VEL. RANGE
respectively PRF
The Velocity Range display is governed by the pulse repetition frequency (PRF)
Exceeding the maximum PRF, the HPRF-Mode is automatically switched on.
12
DYN.
Dynamic Range adjusts the display cutoff of the Doppler analysis waveform.
13
REJECT
Low echo information below the adjusted reject level will not be displayed.
14
CENTER FREQUENCY
It serves for selection of the required transmit frequency.
15
SCALE
To select the displayed measuring unit (in relation to the zero-line).
16
FORMAT
For selection of either one of three formats.
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......