6-4
Section 6-4 - Control Console, Transport Lock
Section 6-4
Control Console, Transport Lock
6-4-1
Control Console
The control console can be rotated 30° to the right.
When rotating the control console grasp it only on the front grip of the user interface.
1.) Push the bolt under the control console forward.
2.) Rotate the console to wanted position.
6-4-2
Transport Lock
There is a locking bolt for locking and unlocking the control console, mounted at the front below the
control console. When preparing the system for transport, the lock has to be engaged in order to secure
the console against uncontrolled rotation. The lock catches in when the console is rotated to its center
0° position.
Figure 6-3 Locking Bolt under Control Console
WARNING
WARNING
Do no
Do not put your hand between the control console and the Main unit when moving the
console to the 0 position: Danger of injuries!
Locking Bolt
for locking and unlocking
the control console
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......