Section 10-7 - Electrical Safety Tests
7.) Follow the test conditions described for respective test points shown in
Table 10-13
.
8.) Keep a record of the results with other hard copies of PM data kept on site.
Table 10-13 Chassis Leakage Current Test Condition
TEST
CONDITION
1
Mounting screw for probe receptacle
2
Wheel support
3
Mounting screw for CRT housing
4
Mounting screw for peripheral plugged into unit
5
Mounting screw for other peripheral powered by unit