® 730 S
ERVICE
M
ANUAL
2-2
Section 2-2 - General Console Requirements
Section 2-2
General Console Requirements
2-2-1
Console Environmental Requirements
2-2-1-1
Cooling
The cooling requirement for the Voluson® 730 is 2500 BTU/hr This figure does not include cooling
needed for lights, people, or other equipment in the room. Each person in the room places an additional
300 BTU/hr. demand on the cooling system.
2-2-1-2
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
2-2-2
Electrical Requirements
NOTE:
GE Medical Systemsrequires a dedicated power and ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before
the system. The dedicated line shall consist of one phase, a neutral (not shared with any other
circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet. Please
note that image artifacts can occur, if at any time within the facility, the ground from the main
facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-2-1
Voluson® 730 Power Requirements
2-2-2-2
Inrush Current
Inrush current is not a factor to consider due to the inrush current limiting properties of the power
supplies.
Table 2-2
Environmental Requirements
Operating
Temperature
Operating
Humidity
Heat Dissipation
Storage
Temperature
Storage Humidity
10 to 40
o
C
(50 to 104
o
F)
30 to 80% rH
non-condensing
2500 BTU pr hour
-10 to 40
o
C
(14 to 104
o
F)
< 90% rH non-
condensing
Table 2-3
Electrical Specifications for Voluson® 730
Voltage
Tolerances
Current
Frequency
100 VAC
±10%
10.10 A
50, 60 Hz (±2%)
115 VAC
±10%
8.80 A
50, 60 Hz (±2%)
130 VAC
±10%
7.80 A
50, 60 Hz (±2%)
230 VAC
±10%
4.40 A
50, 60 Hz (±2%)
240 VAC
±10%
4.20 A
50, 60 Hz (±2%)
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......