Chapter 7 Diagnostics/Troubleshooting
7-3
Section 7-3
Check Points Voltages
7-3-1
How to check power
Figure 7-2 Connectors
Pin 15
+5V
Pin 13
+15V
Pin 11
-15V
ECG
Check above Voltage with DVM to Ground
DVM
Check with DVM (Digital Volt Meter) the 58V/DC
Check above Voltages with DVM to Ground
1
8
9
15
(Backpanel of GEF)
(SCSI Power-Connector Backpanel of GEF)
(Power Connection-Cable from CPN to GEF)
+12V
Pin 1
1
2
3
Check Voltage with DVM to Ground
both Pins 2 and 3 are + 3.3V
Pin 1 is Ground