Chapter 10 Periodic Maintenance
10-7
10-5-2
Functional Checks (See Also Chapter 4)
The functional checks take about 60 minutes to perform. Refer to the system user documentation
whenever necessary.
10-5-2-1
System Checks
NOTE:
* Some software may be considered standard depending upon system model configuration.
Table 10-5 System Functional Checks
Step
Item
Description
1
B-Mode
Verify basic B-Mode (2D) operation. Check the basic system controls that affect this mode
of operation.
2
M-Mode
Verify basic M-Mode operation. Check the basic system controls that affect this mode of
operation.
3
C-Mode
Verify basic C-Mode (Color Flow Mode) operation. Check the basic system controls that
affect this mode of operation.
4
PD-Mode
Verify basic PD-Mode (Power Doppler Mode) operation. Check the basic system controls
that affect this mode of operation.
5
Doppler Modes
Verify basic Doppler Mode operation (PW and CW if available). Check the basic system
controls that affect this mode of operation.
6
3D-Mode
Verify basic 3D-Mode operation. Check the basic system controls that affect this mode of
operation.
7
*Applicable Software
Options
Verify the basic operation of all optional modes such as Real Time 4D, RT_4D Biopsy,
VOCAL, Harmonic Imaging, Sonoview, CRI (Compound Resolution Imaging), TD (Tissue
Doppler), VCI (Volume Contrast Imaging),... etc.
Check the basic system controls that affect each options operation.
8
Keyboard Test
Perform the Keyboard Test Procedure to verify that all keyboard controls are OK.
9
Monitor
Verify basic Monitor display functions.
10
Touch Panel
Verify basic Touch Panel display functions.
11
Measurements
Scan a gray scale phantom and use the measurement controls to verify distance and area
calculation accuracy. Refer to theBasic User Manual, Chapter 13, for measurement
accuracy specifications.
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......