® 730 S
ERVICE
M
ANUAL
7-24
Section 7-8 - Error Messages
In the 3D Image Measure is not allowed
Changed to another format than 3D Fullscreen mode
iSetVideoSource(eVideoIntern) function fails
reboot the system
iSetVideoSource(eVideoIntern) function fails
the system will restart itself by pressing OK
key not in list
restart the system
LP_KoefBlock: SamplePRF too big
restart the system
M_Gain Out Of Range
Press ok and save this user-setting once again
M_Reject Out Of Range
Press ok and save this user-setting once again
MC_BaseLinePos Out Of Range
Press ok and save this user-setting once again
MC_Gain Out Of Range
Press ok and save this user-setting once again
MC-TxPower Out Of Range
Press ok and save this user-setting once again
Memory access violation
restart the system
memory allocation error
restart the system, If after restart not ok then call technical
support
Missing ProbeAcousticUnit, wrong ProbeAcousticUnitID.
restart the system
MotCtrl: No Referenzposition signal!
confirm the message by pressing the OK button,
disconnect and connect the active probe and use it again, if
again the message appears mostly the probe itself has an
defect, so you should contact the service
MotCtrlDrv:bCmdToRS232
restart the system
M-TxPower Out Of Range
Press ok and save this user-setting once again
No CD Writer found
check the SCSI and the Power cable - plug the SCSI cable
off and on and try again. (restart the system)
No disc in drive
insert disk, if fails again reboot and try again (with another
disk)
No Line Memory!
restart the system
No Probe connected on left probe connector (A).
connect the probe
No Probe connected on left probe connector (B).
connect the probe
No Probe connected on left probe connector (C).
connect the probe
OCP connection failed\nPress OK and\nPlease switch
POWER OFF and ON again...
Check the cables from the console, and restart (system must
be switched off)
Old version of flashcomplete!\nFlashcomplete will be
updated!\nPlease wait until confirm message appears!
execute it, restart the system
Overflow
restart the system
Persistence Out Of Range
Press ok and save this user-setting once again
Please plug off and on probe and try again
plug of and on the probe and try again, plug on a different
probe connector.
Error Messages
Actions
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......