Section 4-4 - Functional Checks
4-4-7
Basic Measurements
NOTE:
Different menus are displayed depending on which Mode is selected.
NOTE:
The following instructions assume that you first scan the patient and then press
FREEZE
.
4-4-7-1
Distance and Tissue DepthMeasurements (B- and M-Mode)
1.) Press the
MM
key once.
2.) Touch the appropriate item on the Touch Panel. An active cursor appears.
3.) To position the active cursor at the start point (distance) or the most anterior point (tissue
depth), move the
TRACKBALL
.
4.) To fix the start point, press
SET
(the right or left trackball key). The system fixes the first cursor
and displays a second active caliper.
5.) To position the second active caliper at the end point (distance) or the most posterior point
(tissue depth), move the
TRACKBALL
.
6.) To complete the measurement, press
SET
. The system displays the distance or tissue depth
value in the measurement results window.
Before you complete a measurement:
To toggle between active calipers, press
CHANGE
(upper trackball key).
To erase results, touch the DELETE key on the Touch Panel or press the
CLEAR ALL
key on the
control panel.
NOTE:
The
CHANGE
key alternates the control from one cursor to the other.
NOTE:
To exit Basic measurements, touch the
EXIT
key on the Touch Panel of press the
MM
key on
the control panel.
General remarks to perform Basic measurements:
• By pressing the
MM
key on the control panel the Basic Measurement function is switched on.
• Positioning of measurement marks is done with the
TRACKBALL
.
• Entering and storage of measuring marks is done with
SET
(right or left trackball key).
• To change measuring marks before completion press
CHANGE
(upper trackball key).
• The status bar area shows the current function of the trackball.
• To erase measurement results, touch the DELETE key on the Touch Panel or press the
CLEAR
ALL
key on the control panel.
• To exit from Basic measurements touch the EXIT key on the Touch Panel or press the
MM
key on
the control panel.
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......